Sila koja pokrece dinamizaciju klinova samodinamizirajuceg unutrasnjeg trohanternog fiksatora po Mitkovicu (SIF)
2013
Dynamic trochanteric fractures implants allow fracture fragments to be
compressed. Dynamisation can be realized if the axial pin force overcome
friction force between pin and body of the implant. Examination of sliding
iniciation forces in Mitkovic Selfdinamysible Trochanteric Internal Fixator
(SIF). SIF was attached for angle block in the position with vertical
orientation of pins. The transversal load of 5 kg was connected to pins by a
rope. A dynamometer was used to measure force during the movement of angle
block in up direction. Regression coefficients were a1=4,052 i b1=0,623 for
SIF with 2 sliding screws with diameter of 7mm and a2=4,534 i b2=0,422 for
SIF with 1 screw with diameter of 10 mm. Coefficients of determination were:
r12=0,470 and r22=0,123. Sliding of SIF pins can be achieved for each
analysed body weight of patient (50-130 kg). Early bearing of operated leg is
significant for sliding initiation of SIF sliding screws. [Projekat
Ministarstva nauke Republike Srbije, br. III47107: Virtuelni kostano-zglobni
sistem coveka i njegova primena u pretklinickoj i klinickoj praksi]