Subnanoliter Volume Wall-Jet Cells Combined with Interdigitated Microarray Electrode and Enzyme Modified Planar Microelectrode
2000
Miniaturized wall-jet type flow cells with an active volume of 0.042−15 nL were fabricated for use as highly sensitive electrochemical detectors for capillary electrophoresis/electrochemical detection and small on-line enzyme sensors. The cells consisted of three glass plates and a fused-silica capillary. Two of the plates had microfabricated flow channels and guide trenches for the capillary and working, reference, and counter electrodes. The other plate had a film electrode. When an interdigitated microarray electrode (total area, 66 μm × 64 μm; bandwidth and gap, 2 μm) was installed in the flow cell, the redox cycling enhanced the current at flow rates of less than 100 nL/min even though there were only eight pairs of microbands. A sharp dopamine peak enhanced by the redox cycling was observed when the cell was used for capillary electrophoresis. A square film electrode modified with glutamate oxidase and Os-poly(vinylpyridine) containing HRP was also installed in the flow cell and used to measure neur...
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
27
References
44
Citations
NaN
KQI