Comparison of a new biometer using swept-source optical coherence tomography and a conventional biometer using partial coherence interferometry

2018 
The aim of this study was to compare the axial lengths (ALs) using a new biometer with swept-source optical coherence tomography (Argos) versus ALs using a conventional biometer with partial coherence interferometry (IOL Master, version 5). The ALs in 48 eyes of 48 cataract patients were measured with Argos using refractive indexes that correspond to the particular tissue and with IOL Master using a single refractive index. The eyes were divided into three subgroups by AL length: short-AL group (n = 16), <23.27 mm; intermediate-AL group (n = 16), 23.27–24.03 mm; long-AL group (n = 16), ≥24.04 mm. The ALs (mm) measured with the Argos and IOL Master biometers, respectively, were 22.77 ± 0.43 and 22.74 ± 0.44, 23.63 ± 0.21 and 23.62 ± 0.21, and 26.00 ± 1.61 and 26.05 ± 1.64 in the short-, intermediate-, and long-AL groups, respectively. The mean ALs with the Argos biometer were longer than those with the IOL Master biometer in the short-AL group (P = 0.002) There was no significant difference in the intermediate-AL groups (P = 0.14). In contrast, the mean ALs with the Argos biometer were shorter than those with the IOL Master biometer in the long-AL group (P < 0.001). Differences between the ALs measured with the two biometers were statistically significant in short- and long-AL subgroups. However, the differences might not be clinically significant.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    26
    References
    17
    Citations
    NaN
    KQI
    []