Robotic Mitral Valve Annuloplasty With Double-Arm Nitinol U-Clips

2005 
Abstract Purpose Robotic mitral valve repair increases precision however operative times are longer. Prior studies have indicated that robotic knot tying is time consuming and it is without potential room for improvement. We therefore investigated tissue approximation devices that may shorten operative times. Description A 67-year-old female was approached through a right mini-thoracotomy with the da Vinci Robotic Surgical System (Intuitive Surgical, Sunnyvale, CA). Using 12 nitinol U-clips (Coalescent Surgical, Sunnyvale, CA) an annuloplasty band was placed under robotic guidance. Clip placement and deployment times were recorded and statistical comparisons were assessed to prior suture annuloplasties. Evaluation Clip placement time was 1.3 ± 0.9 (minutes ± standard deviation), statistical comparison with first, most recent, and all prior suture annuloplasties proving no significance. Clip deployment time was 0.5 ± 0.2, whereas knot-tying times and respective statistical comparison for first, most recent, and all prior suture annuloplasties were 2.0 ± 0.7 ( p = 0.003), 1.2 ± 0.4 ( p = 0.0004), and 1.6 ± 0.6 ( p Conclusions U-clips considerably reduce time for annuloplasty over conventional suture and may help reduce operative times as well.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    7
    References
    24
    Citations
    NaN
    KQI
    []