Performance characteristics of the light-initiated chemiluminescent assay for quantitative determination of progesterone

2021 
Background To study the performance of quantitative determination of progesterone by light-initiated chemiluminescent assay (LICA). Methods Clinical samples of serum were used for detection of progesterone by LICA. The precision study was performed according to Clinical and Laboratory Standards Institute (CLSI) EP15-A3, the linear range validation was performed according to CLSI EP06-A, accuracy was evaluated according to CLSI EP9-A3, and the performance of detection capability was confirmed according to CLSI EP17-A2. All data were analyzed using SPSS software. Function regression analysis was performed by OriginPro software. Results The LICA-800 system exhibited low coefficients of variation (CVs) and high reproducibility, and the calculated synthetic CV was 2.16%. The access progesterone assay showed excellent linearity in the assay measuring range (0.37-40 ng/mL) using the polynomial regression method in accordance with CLSI EP06-A. Bias assessment was used to verify accuracy, and the percentage deviation met the quality requirements of the laboratory's allowable deviation of 10.00%. In terms of the detection capability of LICA, the calculated limit of blank (LoB) was 0.046 ng/mL, limit of detection (LoD) was 0.057 ng/mL, and the limit of quantitation (LoQ) value was 0.161 ng/mL. Conclusions The competitive LICA provided a highly sensitive, accurate and precise method for measuring serum progesterone level.