The potential use of ultra-low radiation dose images in digital mammography—a clinical proof-of-concept study in craniocaudal views

2015 
Objective:To estimate the potential of low-dose images in digital mammography by analysing the effect of substantial dose reduction in craniocaudal (CC) views on clinical performance.Methods:At routine mammography, additional CC views were obtained with about 10% of the standard dose. Five radiologists retrospectively read the standard [mediolateral oblique (MLO) + CC] and combination low-dose mammograms (standard MLO + low-dose CC). If present, lesion type, conspicuity and suggested work-up were recorded. Final diagnoses were made by histology or follow up. A t-test or χ2 test was used to compare results.Results:421 cases were included, presenting 5 malignancies, 66 benign lesions and multiple non-specific radiologic features. Using MLO with low-dose CC, all lesions were detected by at least one reader, but altogether less often than with standard mammography (sensitivity, 73.9% vs 81.5%). Missed lesions concerned all types. Lesions detected with both protocols were described similarly (p = 0.084) with c...