Magnetic Particle Imaging (MPI) uses sinusoidal drive fields to excite the magnetic nanoparticles. These time-varying magnetic fields form electric fields within the body, which in turn can cause peripheral nerve stimulation, also known as magnetostimulation. In this work, we propose a design for a human head-size magnetostimulation coil with a Rutherford cable winding. This design achieves 12-fold decrease in the voltages needed to generate a given magnetic field, facilitating the safety of human subject experiments. With electromagnetic simulations, we determine the electric field patterns on a human head model to determine the potential primary locations of magnetostimulation.    Int. J. Mag. Part. Imag. 6(2), Suppl. 1, 2020, Article ID: 2009063, DOI: 10.18416/IJMPI.2020.2009063