Measurement of neutron beam spot distribution using a Micromegas detector for the Back-n white neutron facility at China Spallation Neutron Source.

2019 
The Back-n white neutron beam line, which uses back-streaming white neutrons from the spallation target of China Spallation Neutron Source, is extremely useful for nuclear data measurements. A Micromegas-based neutron detector with two variations is developed to measure the beam spot distribution for this beam line. In this article, the design, fabrication, and characterization of the detector are described. The results of the detector performance tests are presented, which include the relative electron transparency, the gain and the gain uniformity, and the neutron detection capability. The first online test of the detector with the Back-n neutron beam is also presented.