Subharmonic Scattering of SonoVueTM Microbubbles within 10-40 mmHg Overpressures in Vitro
2021
Ultrasound contrast agent microbubbles are considered promising sensors to measure portal vein pressure noninvasively. In this study, we investigated the subharmonic scattering power and optimal incident acoustic pressure of SonoVueTM microbubbles (concentration: 1μL/mL 0.9% NaCl solution) in the ambient pressure range of 10-40 mmHg with 10 mmHg increments at a temperature of 25 °C. The results demonstrated that the subharmonic response of the SonoVueTM microbubbles existed in three stages: the first growth stage (40-300 kPa), saturation (300 kPa-400 kPa), and the second growth stage (400 kPa-540 kPa). In the first growth stage, the subharmonic amplitude increased with ambient pressure. However, while the ambient pressure increased, the subharmonic amplitude decreased in the second growth stage. The best correlation of the subharmonic amplitudes with the ambient pressures was obtained at a high incident acoustic pressure of 520 kPa (sensitivity: 0.15 dB/mmHg, r2 = 0.99, root mean square error = 0.49 mmHg), which indicated that the subharmonic signals in the second growth stage might be suitable for estimating low ambient pressures. The results presented in our study may pave the way for portal vein pressure estimation using SonoVueTM microbubbles as sensors in clinical applications.
Keywords: