A novel approach for quantitation of glucosylceramide in human dried blood spot using LC-MS/MS.

2015 
Background: Glucosylceramide, an efficacy biomarker for Gaucher Type 1 disease, exhibits poor solubility in polar solvents and whole blood which makes it difficult to prepare a homogenous blood standard. Results: We developed a novel method using standard addition approach by spiking a small volume of analyte solution on the surface of prespotted dried blood spot. The whole spots were punched out for subsequent extraction and LC–MS/MS analysis. The assay performance met all validation acceptance criteria. Glucosylceramide concentrations in 50 paired plasma and dry blood spot samples obtained from Gaucher Type 1 patients were tested and the results demonstrated the feasibility of using the DBS method for clinical biomarker monitoring. Conclusion: The new approach greatly improves assay precision and accuracy.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    26
    References
    8
    Citations
    NaN
    KQI
    []