Can Radiocarpal-Spanning Fixation Be Made More Functional by Placing the Wrist in Extension? A Biomechanical Study Under Physiologic Loads

2016 
We investigate whether applying an internal radiocarpal-spanning plate with the wrist in slight extension affects the biomechanical stability of the construct. An unstable distal radius fracture was simulated in 10 cadaveric specimens and immobilized with a radiocarpal-spanning plate holding the wrist in a neutral position. This construct was then physiologically loaded through the wrist flexor and extensor tendons. The resulting motion at the fracture was captured with a displacement sensor. The plate was then extended using an in situ bending technique, placing the wrist in extension, and the experiment was repeated. No statistically significant difference in the biomechanical stability afforded by the radiocarpal-spanning plate was detected with the wrist in extension compared to that in the traditional neutral position. The radiocarpal-spanning plate fixation was more stable when loaded through the extensor tendons. We conclude that immobilizing a distal radius fracture with an internal radiocarpal-sp...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    32
    References
    0
    Citations
    NaN
    KQI
    []