Analysis of Failure Mechanisms in Cemented Short-Stem Femoral Resurfacing Implant

2018 
The effects of metal ion release and wear particle debris in metal-on-metal articulation warrants an investigation of alternative material, like ceramics, as a low-wear bearing couple [1]. Short-stem resurfacing femoral implant, with a stem-tip located at the centre of the femoral head, appears to provide a better physiological load transfer within the femoral head and therefore seems to be a promising alternative to the long-stem design [2]. The objective of this study was to investigate the effect of evolutionary bone adaptation on load transfer and interfacial failure in cemented metallic and ceramic resurfacing implant.Bone geometry and material properties of 3D finite element (FE) models (intact, short-stem metallic and ceramic resurfaced femurs of 44 mm head diameter) were derived from the CT scan data. The FE models consisted of 170352 quadratic tetrahedral elements and 238111 nodes with frictional contact at the implant-cement (μ = 0.3) and stem-bone interfaces (μ = 0.4) and fully bonded cement-bo...