Pre-operative transthoracic real-time three-dimensional echocardiography for a better surgical strategy

2011 
In this study we aimed to evaluate the three-dimensional (3D) transthoracic echocardiography (TTE) in the assessment of cardiac valve morphology. Methods: Bidimensional (2D) and real-time 3D TTE was performed in 104 patients consecutive with cardiac catheterisation, prior to valve surgery. Using surgical findings as the gold standard, 2D and 3D TTE were compared for adequate recognition and accurate detection of morphology. A scoring protocol was used for recognition of the valvular segments (0=inadequate, 1=adequate). Results: Adequate echographic visualization of the valve segments was more frequently obtained by 3D than 2D TTE imaging (731/770 by 3D TTE vs. 693/770 by 2D TTE, p<0.01). The valve leaflets segments were more clearly identified by 3D TTE rather than by 2D TTE (502/531 vs. 471/531, p<0.01). The assessment of commissures was similar by both methods (229/239 vs. 222/239, p=0.09). Total 3D TTE scores
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    33
    References
    0
    Citations
    NaN
    KQI
    []