Evaluation of time to clearance of high‐dose methotrexate using the ARK immunoassay in pediatric hematology/oncology patients: A single‐center experience

2020 
A retrospective, single-center, observational study evaluated the difference in time to clearance of high-dose methotrexate between the fluorescence polarization immunoassay (FPIA) and ARK methotrexate assay in pediatric hematology/oncology patients. The post-ARK immunoassay group had an increase in clearance time of 33 h/cycle and 31% increase in cycles with delayed clearance. On posthoc analysis, use of an adjusted clearance threshold of <0.15 µmol/L post-ARK immunoassay, as opposed to the traditional <0.1 µmol/L threshold, would have similar incidence of delayed clearance. The ARK immunoassay demonstrated a positive bias compared to the FPIA in clinical practice, which led to an institutional policy change.