Swept-source OCT Angiography of the Retinal Vasculature using Intensity Differentiation Based OMAG Algorithms
2015
Background and Objective—To demonstrate the feasibility of using a 1050 nm swept-source OCT (SS-OCT) system to achieve noninvasive retinal vasculature imaging in human eyes. Materials and Methods—Volumetric datasets were acquired using a ZEISS 1 µm SS-OCT prototype that operated at an A-line rate of 100 kHz. A scanning protocol designed to allow for motion contrast processing, referred to as OCT angiography or optical microangiography (OMAG), was used to scan ~3 mm × 3 mm area in the central macular region of the retina within ~4.5 seconds. Intensity differentiation based OMAG algorithm was used to extract 3-D retinal functional microvasculature information. Results—Intensity signal differentiation generated capillary-level resolution en face OMAG images of the retina. The parafoveal capillaries were clearly visible, thereby allowing visualization of the foveal avascular zone (FAZ) in normal subjects.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
14
References
7
Citations
NaN
KQI