Reduced Radiation Dose with Model Based Iterative Reconstruction Coronary Artery Calcium Scoring

2019 
Abstract Assessing coronary artery calcium (CAC) is a valuable tool for individualizing cardiac risk assessment. In CAC scanning, this technical report assesses the use of a true model-based iterative reconstruction algorithm using forward projected model-based iterative reconstruction (“FIRST”) and assess whether FIRST allows for reduced radiation dose CAC scanning on 320-detector row computed tomography (320-CT). Here, 100 consecutive patients prospectively underwent reduced and standard dose scans. For the patients (59 ± 9 years, 61% male) stratified by Agatston categories 0, 1–10, 11–100, 101–400,> 400, agreement between reduced dose with FIRST versus standard dose with FBP was excellent at 81% (95% CI: 73–88%) with kappa 0.74 (95% CI: 0.64–0.85). Median radiation exposure was 75% lower for reduced (0.35 mSv) versus standard dose (1.37 mSv) scans. In conclusion, agreement was excellent for reduced dose with FIRST and standard dose with FBP in 320-detector row CT CAC imaging in well-established categories of cardiovascular risk. These methods make it possible to reduce radiation exposure by 75%.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    40
    References
    7
    Citations
    NaN
    KQI
    []