Analysis of biological and synthetic ribonucleic acids by liquid chromatography-mass spectrometry using monolithic capillary columns
2005
Ion-pair reversed-phase high-performance liquid chromatography (IP-RP-HPLC) has been evaluated as a method for the fractionation and desalting of ribonucleic acids prior to their characterization by electrospray ionization mass spectrometry. Monolithic, poly(styrene−divinylbenzene)-based capillary columns allowed the rapid and highly efficient fractionation of both synthetic and biological ribonucleic acids. The common problem of gas-phase cation adduction that is particularly prevalent in the mass spectrometric analysis of ribonucleic acids was tackled through a combination of chromatographic purification and the addition of ethylenediaminetetraacetic acid to the sample at a concentration of 25 mmol/L shortly before on-line analysis. For RNA molecules ranging in size from 10 to 120 nucleotides, the mass accuracies were typically better than 0.02%, which allowed the characterization and identification of failure sequences and byproducts with high confidence. Following injection of a 500 nL sample onto a 6...