Gel-Like Human Mimicking Phantoms: Realization Procedure, Dielectric Characterization and Experimental Validations on Microwave Wearable Body Sensors

2021 
A simple and low-cost procedure for gel-like time-durable biological phantoms is presented in this work. Easily accessible materials are adopted, which are able to provide a flexible and controllable method to rapidly realize different kind of tissues. The proposed technique is applied to fabricate various tissue-mimicking phantoms, namely skin, muscle, blood and fat. Their effectiveness is first tested by performing dielectric characterization on a wide frequency range, from 500 MHz up to 5 GHz, and validating the measured dielectric parameters (dielectric constant and conductivity) by comparison with reference models in the literature. Then, a multi-layer phantom simulating the human arm is realized, and a wearable body sensor is adopted to prove the perfect agreement of the biometric response achieved in the presence of the fabricated phantom and that provided by a real human arm.