Validation of a novel ultrasound Doppler monitoring device (earlybird) for detection of microvascular circulatory changes

2019 
OBJECTIVE: In this proof-of-concept study we aim to validate a novel ultrasound Doppler monitoring device for evaluating microcirculation (earlybird) against LDF and pulsed Doppler. METHODS: In ten healthy subjects, we measured microcirculatory function at rest and during different autonomic tests (forced respiration, isometric exercise, Valsalva maneuver and cold pressor). Earlybird, LDF and pulsed Doppler were recorded simultaneously. We performed a ZNCC to determine correlation. RESULTS: The curves for earlybird and LDF or pulsed Doppler correlates visually well. Overall median ZNCC 0.87 (interquartile range 0.77 -0.91) between the LDF and earlybird measurements, and 0.90 (0.82 - 0.95) for pulsed Doppler and earlybird. Median ZNCC for baseline and each provocation test for earlybird against LDF and pulsed Doppler were calculated; baseline: LDF 0.87 (0.73 - 0.97) pulsed Doppler 0.91 (0.81 - 0.94), forced respiration: LDF 0.87 (0.28 - 0.90) pulsed Doppler 0.90 (0.85 - 0.96), isometric exercise: LDF 0.82 (0.59 - 0.90) pulsed Doppler 0.87 (0.68 - 0.94), Valsalva maneuver: LDF 0.88 (0.82 - 0.91) pulsed Doppler 0.94 (0.92 - 0.97) and cold pressor: LDF 0.90 (0.85 - 0.95) pulsed Doppler 0.89 (0.65 - 0.94). CONCLUSION: Earlybird records vasoconstrictions in healthy subjects as well as LDF and pulsed Doppler.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    53
    References
    0
    Citations
    NaN
    KQI
    []