Usefulness of Computed Tomography to Predict Mitral Stenosis After Transcatheter Mitral Valve Edge-to-Edge Repair.

2021 
Multidetector computed tomography (MDCT) can provide valuable information for preprocedural planning of transcatheter mitral valve interventions. However, no data exists on pre-MDCT parameters predicting high transmitral pressure gradient (TMPG) post-MitraClip procedure. We analyzed the preprocedural MDCTs of 156 consecutive patients with mitral regurgitation undergoing MitraClip implantation at our institution. The mean TMPG was assessed by periprocedural transesophageal and pre-discharge transthoracic echocardiography. MDCT-derived mitral annulus area (MAA), anterior-posterior (AP) and medial-lateral (ML) mitral annulus diameters, and mitral valve orifice area (MVOA) were smaller in patients with mean TMPG ≥5 mmHg than those with mean TMPG