A complementary dual-modality verification for tumor tracking on a gimbaled linac system

2013 
Background and purpose: For dynamic tracking of moving tumors, robust intra-fraction verification was required, to assure that tumor motion was properly managed during the course of radiotherapy. A dual-modality verification system, consisting of an on-board orthogonal kV and planar MV imaging device, was validated and applied retrospectively to patient data. Methods and materials: Real-time tumor tracking (RTTT) was managed by applying PAN and TILT angular corrections to the therapeutic beam using a gimbaled linac. In this study, orthogonal X-ray imaging and MV EPID fluoroscopy was acquired simultaneously. The tracking beam position was derived from respectively real-time gimbals log files and the detected field outline on EPID. For both imaging modalities, the moving target was localized by detection of an implanted fiducial. The dual-modality tracking verification was validated against a high-precision optical camera in phantom experiments and applied to clinical tracking data from a liver and two lung cancer patients. Results: Both verification modalities showed a high accuracy (<0.3 mm) during validation on phantom. Marker detection on EPID was influenced by low image contrast. For the clinical cases, gimbaled tracking showed a 90th percentile error (E90) of 3.45 (liver), 2.44 (lung A) and 3.40 mm (lung B) based on EPID fluoroscopy and good agreement with XR-log file data by an E90 of 3.13, 1.92 and 3.33 mm, respectively, during beam on. Conclusion: Dual-modality verification was successfully implemented, offering the possibility of detailed reporting on RTTT performance.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    16
    References
    20
    Citations
    NaN
    KQI
    []