Erroneous measurement of the aortic annular diameter using 2-dimensional echocardiography resulting in inappropriate CoreValve size selection: a retrospective comparison with multislice computed tomography.

2014 
Objectives This study sought to assess the differential adherence to transcatheter heart valve (THV)-oversizing principles between transesophageal echocardiography (TEE) and multislice computed tomography (CT) and its impact on the incidence of paravalvular leak (PVL). Background CT has emerged as an alternative to 2-dimensional TEE for THV sizing. Methods In our early experience, TEE-derived aortic annular diameters determined THV size selection. CT datasets originally obtained for vascular screening were retrospectively interrogated to determine CT-derived annular diameters. Annular dimensions and expected THV oversizing were compared between TEE and CT. The incidence of PVL was correlated to TEE- and CT-based oversizing calculations. Results Using TEE-derived annulus measurements, 157 patients underwent CoreValve implantation (23 mm: n = 66; 29 mm: n = 91). The estimated THV oversizing on the basis of TEE was 20.1 ± 8.2%. Retrospective CT analysis yielded larger annular diameters than TEE (p Conclusions Retrospective CT-based annular analysis revealed that CoreValve size selection by TEE was incorrect in 50% of patients. The percentage of oversizing with CT was one-half of that calculated with TEE resulting in the majority of patients receiving a THV that was too small.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    20
    References
    47
    Citations
    NaN
    KQI
    []