Simultaneous determination of levofloxacin, gatifloxacin and moxifloxacin in serum by liquid chromatography with column switching
2004
Abstract Liquid chromatography with a column-switching technique was developed for simultaneous direct quantification of levofloxacin, gatifloxacin and moxifloxacin in human serum. Serum samples were injected on a LiChroCART ® 4-4 pre-column (PC) filled with a LiChrospher ® 100 RP-18, 5 μm where fluoroquinolones (FQs) were purified and concentrated. The FQs were back-flushed from the PC and then separated on a Supelcosil ABZ+ Plus (150 mm × 4.6 mm i.d.) analytical column with a mobile phase containing 10 mM phosphate buffer (pH 2.5), acetonitrile (88:12, v/v) and 2 mM tetrabutyl ammonium bromide. The effects of ion-pair reagents, buffer type, pH and acetonitrile concentrations in the mobile phase on the separation of the three FQs were investigated. Fluorescence detection provided sufficient sensitivity to achieve a quantification limit of 125 ng/ml for levofloxacin and moxifloxacin; 162.5 ng/ml for gatifloxacin with a 5 μl sample size. The on-line process of extraction avoids time-consuming treatment of the samples before injection and run time is shortened. The recovery, selectivity, linearity, precision and accuracy of the method are convenient for pharmacokinetic studies or routine assays.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
13
References
98
Citations
NaN
KQI