Temporal-spatial characteristic evaluation in a dynamic flat-panel detector system

2010 
This report presents the fundamental temporospatial characteristics of a dynamic flat-panel detector (FPD) system. We investigated the relationship between pixel value and X-ray pulse output, and examined reproducibility, dependence on pulse width, tube voltage, and pulse rate. Sequential images were obtained using a direct conversion-type dynamic FPD. The exposure conditions were: 110 kV, 80 mA, 6.3 ms, 7.5 fps, source-to-image distance (SID) 1.5 m. X-ray pulse output was measured using a dosimetry system with a sampling interval of 70 μs, to determine temporal changes in each X-ray pulse output. Temporal changes in pixel value were measured in the obtained images, and the relationship between pixel value and X-ray pulse output was examined. Reproducibility was assessed by comparing the results in two sequential images obtained under the same exposure conditions. Moreover, the relationships and properties were evaluated by changing the pulse width (12 ms and 25 ms), tube voltage (80 kV, 90 kV, and 100 kV), and pulse rate (3.75 fps and 15 fps). The results showed a good correlation between the X-ray pulse output and pixel values. Fluctuation of the pixel value measured in sequential images is thought to be mainly due to changes in X-ray pulse output, and is not caused by FPD.