Aortic dimensions by multi-detector computed tomography vs. echocardiography

2016 
Abstract Objective Clinical follow-up of aortic dimensions is performed interchangeably by multi-detector computed tomography (MDCT) and by cardiac echocardiography (ECHO). This study assesses the relationship between measurements of the aortic diameter by MDCT and ECHO at various predetermined locations using several methods. Methods The aortic diameter was measured at 6 locations between the aortic annulus and the aortic arch in 49 patients who underwent both MDCT and ECHO. Measurements were performed by three methods: internal-to-internal edge (INT), external-to-internal edge (MIX), and external-to-external edge (EXT). Measurements by MDCT and ECHO were made by an experienced radiologist and cardiologist, respectively, both blinded to results and images from the other modality. Results The average aortic diameter at all locations was significantly different between the MDCT and ECHO by all three methods (INT: 30.0 ± 5.8 mm vs. 27.8 ± 5.9 mm; MIX: 31.5 ± 5.8 mm vs. 30.8 ± 5.8 mm; EXT: 32.9 ± 6.6 mm vs. 33.8 ± 6.5 mm, p p p Conclusions There is considerable variability between MDCT and ECHO measurements of the ascending aorta. Measuring the aortic diameter by the MIX provides the closest measurements and is advised for long-term follow-up.