Comparative TQT analysis with three fully-automated platforms: Comparison to core laboratory semi-automated results☆

2013 
Abstract Background and Purpose Technological advances in machine-read QT measurement now enable detailed and precise cardiac repolarization assessments. This study assessed the applicability of three state-of-art ECG measurement applications to provide reliable continuous analyses from data obtained in a positive thorough QT study previously characterized with sparse semi-automated measurements performed by an ECG core laboratory. Methods Continuous RR, QT, QTc measurements, and individual QT/RR relationships and their associated intra- and inter-subject variability were derived in parallel with BioQT, Ponemah PRO, and WinAtrec analysis software. Results Despite slight vendor-specific differences in measurement variability and QTc, all machine-read methods demonstrated requisite assay sensitivity and yielded similar conclusions in accordance with SA analysis. Conclusions Three commercially available ECG analytical software applications reliably detected the drug-induced QT prolonging effects and replicated the SA core-laboratory conclusions, with greatly improved temporal resolution and reduced analytical costs. With broader experience, these data suggest that current SA methodologies could be effectively replaced by fully automated ECG analysis.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    23
    References
    1
    Citations
    NaN
    KQI
    []