Inflation and deflation timing of the AutoCAT 2 WAVE intra-aortic balloon pump using the autoPilot mode in a clinical setting

2012 
The primary goal of this observational clinical study was to register the occurrence of incorrect inflation and deflation timing of an intra-aortic balloon pump in autoPilot mode. The secondary goal was to identify possible causes of incorrect timing. During IABP assistance of 60 patients, every four hours a strip was printed with the IABP frequency set to 1:2. Strips were examined for timing discrepancies beyond 40 ms from the dicrotic notch (inflation) and the end of the diastolic phase (deflation). In this way, 320 printed strips were examined. A total of 52 strips (16%) showed incorrect timing. On 24 of these strips, the incorrect timing was called incidental, as it showed on only one or a few beats. The other 28 cases of erroneous timing were called consistent, as more than 50% of the beats on the strip showed incorrect timing. We observed arrhythmia in 69% of all cases of incorrect timing. When timing was correct, arrhythmia was found on 13 (5%) of 268 strips. A poor quality electrocardiograph (ECG)...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    17
    References
    2
    Citations
    NaN
    KQI
    []