Biomechanical analysis of the cephalomedullary nail versus the trochanteric stabilizing plate for unstable intertrochanteric femur fractures.

2016 
Unstable intertrochanteric fractures are commonly treated with a cephalomedullary nail due to high failure rates with a sliding hip screw. The Omega3 Trochanteric Stabilizing Plate is a relatively new device that functions like a modified sliding hip screw with a proximal extension; however, its mechanical properties have not been evaluated. This study biomechanically compared a cephalomedullary nail, that is, Gamma3 Nail against the Omega3 plate. Unstable intertrochanteric fractures were created in 24 artificial femurs. Experimental groups were as follows: Nail (i.e. Gamma3 Nail) (n = 8), Plate A (i.e. Omega3 plate with four distal non-locking screws and no proximal locking screws) (n = 8), Plate B (i.e. Plate A plus five proximal locking screws) (n = 8), Plate C (i.e. Omega3 plate with three distal locking screws and no proximal locking screws) (n = 8), and Plate D (i.e. Plate C plus five proximal locking screws) (n = 8). All specimens were stiffness tested, while the Nail and Plate D groups were also s...