Investigation of a new design for zirconia dental implants

2007 
Abstract Objective To assess the biomechanical properties of a new design configuration for zirconia dental implants. Methods The new design has a cylindrical shape that is partially hollow and porous in the bottom, which permits the implants to be locked into the alveolar bone over time. It also utilizes bioactive glass coatings to increase adhesion to surrounding bone structure. Samples of the new design were fabricated in the laboratory and their material strength, hardness, and fracture toughness were evaluated. In addition, biocompatibility of the new design was evaluated through testing in dogs. Results Results of mechanical tests indicate that structural properties of the new design exceed the usual requirements for implants. Moreover, animal tests suggest that there is appreciable improvement in lock-in strength and osteointegration. Conclusion The new design configuration is biomechanically feasible and further research is warranted to improve the design for human use.
    • Correction