Correlation Between Average Segment Width and Gamma Passing Rate as a Function of MLC Position Error in Volumetric Modulated Arc Therapy.

2021 
OBJECTIVE This study analyzed the correlation between the average segment width (ASW) and gamma passing rate according to the multi-leaf collimator (MLC) position error. METHOD To evaluate the changes in the gamma passing rate according to the MLC position error, 21 volumetric modulated arc therapy (VMAT) plans were generated using pelvic lymph node metastatic prostate cancer patient's data which is sensitive to MLC position errors as they involve several long, narrow, irregular fields. The ASW for each VMAT plan was calculated using our own code developed using Visual Basic for Applications (VBA). The gamma passing rate of the VMAT plan according to the MLC position error was evaluated using ArcCHECK (Sun Nuclear, Melbourne, FL, USA) while inducing symmetric MLC position errors in 0.25 mm intervals from -1 mm to +1 mm in the infinity medical linear accelerator (Elekta AB, Stockholm, Sweden). Finally, we examined the correlation between the change in the passing rate (γgradient) due to the MLC position error and the ASW in VMAT through linear regression analysis using the least squares method. RESULTS The ASW and γgradient were found to have a linear correlation according to the MLC position error, and the coefficient of determination was 0.88. For a 1 mm position error of MLC in VMAT, the gamma passing rate improved by approximately 11.9% as the ASW increased by 10 mm. CONCLUSION These results are expected to be employed as guidelines to minimize the dose uncertainty due to MLC position error in VMAT.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    11
    References
    0
    Citations
    NaN
    KQI
    []