Reconstructed helical CT scans: improvement in z-axis resolution compared with overlapped and nonoverlapped conventional CT scans.

1995 
This in vitro study was designed to assess the z-axis resolving capabilities of reconstructed helical CT scans obtained with various imaging parameters versus those of conventional CT scans and the effect of decreasing slice index on the z-axis resolution of helical CT.A z-axis line-pair phantom was imaged using conventional nonoverlapped CT scans, conventional CT scans that overlapped by 50%, and helical CT scans with pitches of 1.0 and 1.5. All helical images were reconstructed at comparable slice indexes (image indexes of 2.0, 1.0, and 0.5 mm for pitch = 1.0, and image indexes of 3.0, 1.5, and 0.75 mm for pitch = 1.5). Midline coronal and sagittal reconstructed images were obtained to allow standardized visualization of line pairs. The reconstructed images were reviewed separately by 10 radiologists.The overall z-axis resolution of reconstructed helical CT scans equaled or exceeded that of nonoverlapped conventional CT scans in all cases and equaled that of 50% overlapped conventional CT scans in 75% o...