Pre- and Postoperative Accuracy and Safety of a Real-Time Continuous Glucose Monitoring System in Cardiac Surgical Patients: A Randomized Pilot Study

2009 
Abstract Background: Our objective was to evaluate the accuracy and safety of a real-time (RT) continuous glucose monitoring system (CGMS) in patients before and after cardiothoracic surgery and to investigate whether activation of the alarm function of the RT-CGMS had an effect on glucose control. Methods: Patients scheduled for elective cardiothoracic procedures, without a history of insulin-requiring diabetes, were perioperatively monitored with RT-CGMS for 72 h and were randomized into two groups: with or without the alarm function (set at 4 and 10 mmol/L) of the device activated. Sensor values were compared with capillary, arterial, and venous blood glucose values. Percentages of time spent in various glucose ranges were compared between groups. Results: There were no adverse effects of the RT-CGMS. Of the 1,001 sensor value comparisons with capillary or arterial measurements, 96.6% fell within Clarke Error Grid zones A and B, with relative absolute differences ranging from 15% (preoperative period) ...