Analysis of 30 Breast Implant Rupture Cases

2005 
Breast implants used for augmentation mammoplasty or breast reconstruction could rupture from various causes such as trauma or spontaneous failure. The objectives of this study were to investigate the relationships between the causes of implant rupture and the degree of capsular contracture, and then to evaluate the relative efficacies of specific signs on magnetic resonance imaging (MRI) known to be beneficial for diagnosing the rupture. A retrospective review identified patients with prosthetic implant rupture or impending rupture treated by the senior author. The 30 cases of implant rupture available for review were classified into two groups: intracapsular and extracapsular ruptures. The 30 cases of breast implant ruptures were analyzed with respect to the clinical symptoms and signs, the causes of rupture, the degree of capsular contracture, and therapeutic plans. Among the 30 cases, 14 patients who had undergone MRI during the diagnostic period were analyzed with respect to the relationships between MRI readings and operative findings. Spontaneous rupture of membranes was most common (80%), followed by failure because of trauma (7%) and valve or implant base (4%). The symptoms during implant rupture were contour deformity, palpated masslike lesions, pain, and focal inflammation. According to the analysis of specific MRI signs, the sensitivity and specificity of the linguine sign were 87% and 100%, respectively, for intracapsular rupture. For extracapsular rupture, the sensitivity and specificity of the linguine sign were, respectively, 67% and 75%. The sensitivity and specificity of the rat-tail sign and tear drop sign were 14% and 50%, respectively.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    12
    References
    26
    Citations
    NaN
    KQI
    []