Image quality of direct conversion detectors for mammography and radiography: a theoretical comparison
2003
Direct conversion detectors have the potential to provide very high resolution and high detective quantum efficiency (DQE). Selection of a material that is appropriate for the task is dictated by the material properties. A linear cascaded systems analysis of DQE is used to predict the performance of several detector materials such as amorphous Se, CdZnTe, and PbI 2 . A model is used to predict the spatial frequency-dependent DQE ( f ) for each material. This model includes: (1) x-ray absorption, (2) K fluorescence, (3) conversion gain, and (4) incomplete charge collection. A depth-dependent approach is used to account for gain variations and charge transport characteristics that change throughout the detector. In the model a parallel cascade, and non-elementary stages are used to model the effect of K-fluorescence reabsorption followed by incomplete charge collection. The DQE ( f ) is determined across an x-ray energy range of 10 to 100 keV for each material under typical bias conditions ranging from 0.1 V/μm to 10 V/μm. K-fluorescence escape and reabsorption blurring can cause marked reductions in the DQE ( f ). It is further reduced by incomplete charge collection which can theoretically decrease the DQE ( f ) by as much as 50% in extreme situations. This model will help determine key factors that will influence material selection for direct conversion x-ray systems.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
0
References
5
Citations
NaN
KQI