True beam commissioning experience at Nordland Hospital Trust, Norway

2016 
To evaluate the measured of all photon beam data of first Varian True Beam version 2.0 slim model, recently commissioned at Nordland Hospital Trust, Bodo. To compare and evaluate the possibility of beam matching with the Clinac2300, for the energies of 6MV and 15 MV. Materials/Methods: Measurements of PDD, OAR, and Output factors were realized with the IBA Blue-phantom with different detectors and evaluated between them for all photon energies: 6MV, 15MV, 6MV FFF and 10MV FFF. The ionization chambers used were Pin Point CC01, CC04, Semiflex CC13 and photon diode by Iba dosimetry. The data were processed using Beizer algorithm with a resolution of 1 mm. The measured depth dose curves, diagonals, OAR, and output factors were imported into Eclipse in order to calculate beam data for the anisotropic analytical algorithm (AAA version 10.0.28) for both the dataset measured with CC04 and CC13 and compared. The model head of 23EX was selected as the most near model to True Beam as a restriction of our version of ...