High-Tc Superconductor Detection Coils for a Magnetic Resonance System
2002
Considering that in a magnetic resonance system, if the detection coil contributes dominantly to the system noise, the performance of the whole system can certainly be improved by switching to a detection coil made of a high-temperature superconductor, and using YBa2Cu3O7 thin films on 25?25?mm2 LaAlO3 substrates, we have prepared two kinds of detection coils: single-coil and two-coil. Encouragingly, their quality factors are measured to be Q>2500 for two-coil (at 22.566?MHz and 77?K) and Q>5500 for single-coil (at 92.3?MHz and 77?K), respectively. Here, we describe the details of the design, fabrication and testing of the coils.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
4
References
2
Citations
NaN
KQI