Evaluation of High Flow Local Extraction on control of the aerosol plume in an operating theatre

2021 
BackgroundEngineering controls are a necessity for minimising aerosol transmission of SARS-CoV-2, yet so far, little attention has been given to such interventions. High flow local extraction (HFLE) is a standard in other industries that deal with airborne contaminants. ObjectiveThis study aims to provide a quantitative evaluation of an HFLE concept feasible to implement in most real clinical settings. DesignA unique combined experimental model of Laser sheet illumination videography paired with continuous nanoparticle counts was used to quantitatively assess the impact of HFLE in an operating theatre. Propylene Glycol was aerosolised via a customised physiological lung simulator and dispersion was measured in 3 dimensions. Cumulative probability heat maps were generated to describe aerosol behaviour. Continuous particle counts were made at 15 locations throughout the room to validate laser assessments. ResultsHigh flow local extraction reduced dispersion of simulated exhaled aerosols to undetectable levels. With the HFLE in operation and optimally positioned, the aerosol plume was tightly controlled. Particle counts remained at baseline when HFLE was active. HFLE becomes less effective when positioned at increasing distance from the mouth. Aerosol plume behaviour in the absence of HFLE was highly variable and unpredictable. ConclusionsThis analysis demonstrates great potential for HFLE to have a significant impact in reducing aerosol transmission. Simple HFLE devices can be easily engineered and could be widely deployed without impacting on the safe delivery of care.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    35
    References
    0
    Citations
    NaN
    KQI
    []