Rapid in vivo whole body composition of rats using cone beam μCT
2010
Precise, noninvasive analysis and quantification of in vivo body composition is essential for research involving longitudinal, small-animal disease models. We investigated the feasibility and precision of a rapid, flat-panel μCT scanner to report whole body adipose tissue volume (ATV), lean tissue volume (LTV), skeletal tissue volume (STV), and bone mineral content (BMC) in 25 postmortem female and 52 live male Sprague-Dawley rats. μCT images, acquired in three 90-mm segments and reconstructed with 308 μm of isotropic voxel spacing, formed contiguous image volumes of each entire rat specimen. Three signal-intensity thresholds (determined to be −186, 5, and 155 HU) were used to classify each voxel as adipose, lean, or skeletal tissue, respectively. Tissue masses from the volume fractions of ATV, LTV, and STV were calculated from assumed tissue densities of 0.95, 1.05, and 1.92 g/cm−3, respectively. A CT-derived total mass was calculated for each rat and compared with the gravimetrically measured mass, whic...