Development of a 6 LiF/ZnS-based neutron multiplicity counter
2015
Neutron multiplicity counters are used in safeguards to provide rapid assay of samples which contain an unknown amount of plutonium in a potentially unknown configuration. A project at PNNL is using regular and nickel-quenched 6 LiF/ZnS neutron-scintillator sheets and wavelength shifting plastic for light pipes in place of 3 He. A combination of laboratory and modeling work predicts a LiF/ZnS-based system to be able to match or exceed the performance of the best 3 He-based systems available. Also, the Ni-quenched material is expected to allow for improved neutron/gamma-ray discrimination at twice the event rate relative to the non-Ni-quenched LiF/ZnS. A new system based on the LiF/ZnS material is under construction and components are being used to optimize the detection efficiency and neutron/gamma-ray discrimination properties. Components of the new system are partially constructed and undergoing performance testing utilizing high-speed digitizers with field programmable gate arrays to perform the neutron/gamma-ray discrimination. The expected performance of the full-scale system is expected to be nearly the same as for 3 He-based systems and is due for completion in 2016.