Evaluation of reducing temporal resolution on the accuracy of aortic pulse wave velocity assessment from velocity encoded MRI

2014 
Background Aortic pulse wave velocity (PWV), the propagation speed of blood flow velocity waves through the aorta, is a marker of aortic stiffness with prognostic value in various diseases with vascular expression. One-directional throughplane velocity-encoded (VE) MRI, planned perpendicular to the ascending aorta and additionally transecting the proximal descending aorta (Figure 1), is a validated method for assessing PWV over the aortic arch. However, the effect of the temporal resolution (Tres) of VE MRI on the accuracy of PWV assessment has not yet been established. Therefore, the aim of this study was to evaluate the effect of reducing Tres on the accuracy of aortic PWV and how this relates to physiological variation. Methods Five patients referred for cardiac MRI and ten healthy volunteers within similar age range (mean age 32 ± 14 years) were prospectively included. PWV was assessed from velocity mapping using VE MRI on 3T MRI (Ingenia, Philips) with velocity sensitivity of 150 cm/s. Reference PWV (PWVref) was achieved from VE MRI with maximal number of reconstructed phases (Tres = 5 ms). The effect of temporal Tres on PWV was evaluated by temporal downsampling (50% [Tres = 10 ms], 67% [Tres = 15 ms], 75% [Tres = 20 ms], and 80% [Tres = 25 ms]), first by reducing phases during repeated reconstruction of the original acquired high-temporal data. Next, downsampling was performed by data removal as well as by data