Intracoronary Stenting and Angiographic Results

2001 
Background—Increased thrombogenicity and smooth muscle cell proliferative response induced by the metal struts compromise the advantages of coronary stenting. The objective of this randomized, multicenter study was to assess whether a reduced strut thickness of coronary stents is associated with improved follow-up angiographic and clinical results. Methods and Results—A total of 651 patients with coronary lesions situated in native vessels >2.8 mm in diameter were randomly assigned to receive 1 of 2 commercially available stents of comparable design but different thickness: 326 patients to the thin-strut stent (strut thickness of 50 μm) and 325 patients to the thick-strut stent (strut thickness of 140 μm). The primary end point was the angiographic restenosis (≥50% diameter stenosis at follow-up angiography). Secondary end points were the incidence of reinterventions due to restenosis-induced ischemia and the combined rate of death and myocardial infarctions at 1 year. The incidence of angiographic resten...