Long‐Term Outcomes and Durability of the Mitroflow Aortic Bioprosthesis

2016 
Background This study aims to determine the incidence and causes of structural valve deterioration (SVD) among all models of Mitroflow bioprostheses (A12/LX/DL), as well as to define their long-term clinical and hemodynamic performance. Methods and Results We retrospectively reviewed a series of 1023 patients who underwent aortic valve replacement with Mitroflow bioprostheses between 2001 and 2014. A small aortic root was found in 22.4% of patients. There were two cases of severe patient-prosthesis mismatch. Only 31 patients developed SVD. The rate of incidence was 8.1 cases per 1000 patient-years. Cumulative incidence of SVD was 1.4% and 3% at five and 10 years, respectively. Freedom from SVD was 97.4% and 88.2% at five and 10 years, respectively. Anticalcification phospholipid reduction treatment (PRT) for model DL was a protective factor for SVD. Multivariable analysis confirmed age <70 years and use of 19 mm valve as independent predictors of SVD. Cumulative survival was 76.6% at five years and 42.3% at 10 years (mean follow-up 3.8 ± 3.1 years). In multivariable analysis, neither the use of small aortic prosthesis (p = 0.18) nor the occurrence of SVD (p = 0.85) was found to be independent predictors of long-term survival. Conclusions Mitroflow valves demonstrate an acceptable rate of SVD and satisfactory long-term hemodynamic performance, particularly in patients with small aortic roots, age >70 years, and cases with severe left ventricular hypertrophy. PRT might contribute to improved long-term durability.