Analytical validation of an HPLC assay for urinary albumin
2006
Abstract Background Microalbuminuria is the earliest clinical finding for renal disease. Diabetic individuals often produce modified forms of albumin, perhaps due to impaired lysosomal processing, that are undetectable by common immunoassays but accurately measured by HPLC. Methods We evaluated the performance of a commercially available, FDA-approved HPLC assay (AusAm Biotechnologies, NY) and compare results to our immunoturbidimetric assay (ITA, Beckman-Coulter, CA) using random urine specimens from 32 nondiabetic and 60 type 1 and 2 diabetic subjects. Results The HPLC assay was linear to 963 mg/l with a limit of detection of 6.1 mg/l. Within-run and between-run precision was Conclusions Prospective studies are necessary to determine whether the HPLC assay can effectively detect microalbuminuria earlier than current assays without a concomitant increase in the false positive rate.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
16
References
30
Citations
NaN
KQI