Abstract A high-count-rate ionization chamber design with position-sensitivity has been developed and deployed at several accelerator facilities. Counting rates of ≥ 500 kHz with good Z -separation (up to 5% energy resolution) for particle identification have been demonstrated in a series of commissioning experiments. A position-sensitive capability, with a resolution of 3 mm, has been implemented for the first time to record position information and suppress pileup. The design and performance of the detectors are described.