Influence of pacing site characteristics on response to cardiac resynchronization therapy.

2013 
Background—Transmural scar occupying left ventricular (LV) pacing regions has been associated with reduced response to cardiac resynchronization therapy (CRT). However, spatial influences of lead tip delivery relative to scar at both pacing sites remain poorly explored. This study evaluated scar distribution relative to LV and right ventricular (RV) lead tip placement through coregistration of late gadolinium enhancement MRI and cardiac computed tomographic (CT) findings. Influences on CRT response were assessed by serial echocardiography. Methods and Results—Sixty patients receiving CRT underwent preimplant late gadolinium enhancement MRI, postimplant cardiac CT, and serial echocardiography. Blinded segmental evaluations of mechanical delay, percentage scar burden, and lead tip location were performed. Response to CRT was defined as a reduction in LV end-systolic volume ≥15% at 6 months. The mean age and LV ejection fraction were 64±9 years and 25±7%, respectively. Mean scar volume was higher among CRT n...