Accuracy of the new voltage standard using Josephson junctions cooled to 77 K

2016 
A practical dc voltage standard with output voltages Ł/hts from 0.1 V to 10 V using high temperature superconductor (HTS) arrays of Josephson junctions cooled to 77 K was characterized. HTS arrays with increased current margins and the voltage divider with a high stability of the divider's coefficients enable stable and reproducible automated calibration of the new HTS standard. The direct comparison of U HTS against the voltage generated on an array of niobium Josephson junctions cooled to 4.2 K reveals the agreement of voltages at 1 V level with a Type A uncertainty equal to a few parts in 10 8 .
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    4
    References
    4
    Citations
    NaN
    KQI
    []