Calibration of the linear response range of x-ray imaging plates and their reader based on image grayscale values.

2017 
X-ray imaging plates are one of the most important X-ray imaging detectors and are widely used in inertial-confinement fusion experiments. However, their linear response range, which is the foundation of their quantitative data analysis, has not been sufficiently deeply investigated. In this work, we develop an X-ray fluorescer calibration system and carefully explore the linear response range of X-ray imaging plates. For the first time, nearly the entire grayscale range of the X-ray imaging plate linear response—7819–64 879 in the range of 0–65 535—has been observed. Further, we discuss the uncertainties involved in the calibration process. This work demonstrates the excellent linear response qualities of X-ray imaging plates and provides a significant foundation for expanding their quantitative applied range.