Twelve-month angiographic and clinical outcomes of the XINSORB bioresorbable sirolimus-eluting scaffold and a metallic stent in patients with coronary artery disease

2019 
Abstract Background Recent studies showed bioresorbable scaffold (BRS) increased risks of late target lesion failure (TLF) and thrombosis. XINSORB scaffold is a poly-L-lactic acid based BRS. Methods The study included randomization and registry parts. Eligible patients with one or two de novo lesions were randomly 1:1 assigned to XINSORB scaffold and sirolimus-eluting stent (SES) in randomization part. These patients were clinically and angiographically assessed. In registry part, patients were treated with XINSORB scaffold only and were clinically assessed. The primary endpoint was in-segment late luminal loss (LLL) at 12-month in randomization part. The secondary endpoint was 12-month TLF in all XINSORB-treated patients. Results Total 395 and 798 patients were enrolled in randomization and registry part, respectively. Device success was 98.0% (1069/1091) in all XINSORB-treated and 100% (221/221) in SES-treated lesions. The primary endpoint of in-segment LLL at 12-month was 0.19 ± 0.32 mm in XINSORB and 0.31 ± 0.41 mm in SES (P = 0.003), which met the noninferior margin of 0.195 mm (95% CI: −0.20, −0.04, P  Conclusions In the multicenter clinical trial, XINSORB BRS was noninferior to sirolimus-eluting stent for the primary endpoint of in-segment LLL at 12-month in patients with simple and moderate complex de novo coronary lesions. TLF at 12-month was low and comparable.
    • Correction