Long-term electrode impedance changes and failure prevalence in cochlear implants

2015 
AbstractObjective: This study assessed the prevalence of electrode failures and electrode impedance measures in Nucleus cochlear implants around initial activation (an average of 16 days after surgery) and after 8 to 12 years of device use. Design: Retrospective data from the Melbourne Cochlear Implant Clinic was collated and analysed. Study sample: Included in this study were 232 adults, all of whom were implanted at the clinic between March 1998 and August 2005. Results: Overall 0.5% of electrodes failed over the entire test period, with 5.6% of devices showing one or more electrode failure. The majority of these failures were recorded by initial activation. The numbers of electrode failures have decreased over time with array type, such that no failures were recorded with the currently available Contour Advance array. Array type was shown to affect electrode impedance at both time points, with the Contour and Contour Advance arrays having significantly higher absolute values than the Banded array. Howe...