Liver Cancer Detection by a Simple, Inexpensive and Effective Immunosensor with Zinc Oxide Nanoparticles

2015 
Regular monitoring of blood α-fetoprotein (AFP) and/or carcino-embryonic antigen (CEA) levels is important for the routine screening of liver cancer. However, AFP and CEA have a much lower specificity than des-γ-carboxyprothrombin (DCP) to detect liver cancer. Therefore, the study reported here was designed, to develop a screen-printed DCP immunosensor incorporating zinc oxide nanoparticles, for accurate determination of DCP. The designed immunosensor shows low detection limits for the detection of DCP: 0.440 ng/mL (based on impedance measurement), 0.081 ng/mL (based on real part of impedance measurement) and 0.078 ng/mL (based on imaginary part of impedance measurement), within the range of 3.125 ng/mL to 2000 ng/mL. In addition, there was little interference to DCP determination by molecules such as Na+, K+, Ca2+, Cl−, glucose, urea, and uric acid. It is therefore concluded that the DCP immunosensor developed and reported here is simple, inexpensive and effective, and shows promise in the rapid screening of early-stage liver cancer at home with a point-of-care approach.